Keithley 2410 SourceMeter SMU 儀器 數字源表系列
聯系我們詳情
電壓精度(本地或遠端感測)
型號 |
量程 |
設置分辨率 |
電源
1 |
缺省測量分辨率 |
測量 2, 3, 4 準確度 (1 年)23°C ±5°C±(% rdg. + 電壓) |
輸出轉換速率 (±30%) |
源/阱極限 |
2400, 2401 |
200.000 mV |
5 μV |
0.02% + 600 μV |
1 μV |
0.012% + 300 μV |
±21 V @ ±1.05 A ±210 V @ ±105 mA* |
|
2.00000 V |
50 μV |
0.02% + 600 μV |
10 μV |
0.012% + 300 μV |
|
||
20.0000 V |
500 μV |
0.02% + 2.4 mV |
100 μV |
0.015% + 1.5 mV |
0.08 V/μs |
||
200.000 V* |
5 mV |
0.02% + 24 mV |
1 mV |
0.015% + 10 mV |
0.5 V/μs |
||
2410 |
200.000 mV |
5 μV |
0.02% + 600 μV |
1 μV |
0.012% + 300 μV |
±21 V @ ±1.05 A ±1100 V @ ±21 mA |
|
2.00000 V |
50 μV |
0.02% + 600 μV |
10 μV |
0.012% + 300 μV |
|
||
20.0000 V |
500 μV |
0.02% + 2.4 mV |
100 μV |
0.015% + 1 mV |
0.15 V/μs |
||
1000.00 V |
50 mV |
0.02% + 100 mV |
10 mV |
0.015% + 50 mV |
0.5 V/μs |
||
2420 |
200.000 mV |
5 μV |
0.02% + 600 μV |
1 μV |
0.012% + 300 μV |
±21 V @ ±3.15 A ±63 V @ ±1.05 A |
|
2.00000 V |
50 μV |
0.02% + 600 μV |
10 μV |
0.012% + 300 μV |
|
||
20.0000 V |
500 μV |
0.02% + 2.4 mV |
100 μV |
0.015% + 1 mV |
0.08 V/μs |
||
60.0000 V |
1.5 mV |
0.02% + 7.2 mV |
1 mV |
0.015% + 3 mV |
0.14 V/μs |
||
2440 |
200.000 mV |
5 μV |
0.02% + 600 μV |
1 μV |
0.012% + 300 μV |
±10.5 V @ ±5.25 A ±42 V @ ±1.05 A |
|
2.00000 V |
50 μV |
0.02% + 600 μV |
10 μV |
0.012% + 300 μV |
|
||
10.0000 V |
500 μV |
0.02% + 1.2 mV |
100 μV |
0.015% + 750 μV |
0.08 V/μs |
||
40.0000 V |
5 mV |
0.02% + 4.8 mV |
1 mV |
0.015% + 3 mV |
0.25 V/μs |
||
*Not available on 2401. |
NOTES
- 溫度系數(0°C ~18°C & 28°C ~50°C):±(0.15×規定的準確度)/°C。
- 電壓調整率:線路調整率:量程的0.01%。負載調整率:量程的0.01%+ 100μV。
- 過壓保護:用戶可選值,5%允許誤差。工廠缺省=none。
- 電流極限/鉗位:雙極性電流極限(嵌位)設為一個值。量程的0.1%(最小值)。
- 過沖:典型值<0.1%(滿量程步長,電阻性負載,10mA量程)。
源指標附加說明(適用所有型號)
瞬時響應時間:當負載發生階躍變化時,輸出信號恢復到規定值的時間 為30μs(最低要求)。
命令處理時間:接收到:SOURce:VOLTage|CURRent命令后,輸出端開始變化所需的最長時間。自動量程開:10ms。自動量程關:7ms。
輸出穩定時間:命令處理和輸出轉換之后,達到最終值0.1%的所需時間。100μs(典型值)。阻性負載。10μA至100mA量程。
直流浮地電壓:輸出端可以從機殼地電平上浮至±250VDC(2440型為±40VDC)。
遠端感測:每條負載線的壓降達1V。
嵌制準確度:在基本指標上量程增加0.3%并且讀數增加±0.02%。
過溫保護:如果內部檢測到溫度過載就將部件置為待機模式。
量程變化過沖:過沖至全阻性100kΩ負載,10Hz~1MHz帶寬,相鄰
量程:100mV(典型值),除了20V/200V(20V/60V,對于2420),20V/100V對于2425和2430,量程邊界和和2440。
最小嵌位值:量程的0.1%。
附加脈沖模式源指標(僅對2430和2430-C)
最大占空比:8%,硬件限制,僅10A量程。所有其它量程為100%。
最大脈沖寬度:5ms從90%上升沿至90%下降沿,2.5ms 10A量程。
最小脈沖寬度:150μs。
最小脈沖分辨率:50μs(典型值),70μs(最大值)由系統抖動限制。
源精度:由穩定時間和源量程指標決定。
輸出穩定時間0.1%:
800μs(典型值),電流源=10A至10Ω,被電壓轉換速率限制。
500μs(典型值),電流源=10A至1Ω,被電壓轉換速率限制。
輸出轉換速率:
電壓(10Ω負載):0.25V/μs±30%在100V量程。0.08V/μs±30%在20V量程、10A量程。
電流(0Ω負載):0.25A/μs±30%在100V量程。0.08A/μs±30%在20V量程、10A量程。
注解:
1.僅2400、2410:指標對連續輸出電流低于105mA有效。對高于105mA連續工作超過
1分鐘,精度降低10%/35mA(高于105mA)。
2.速度=正常(1PLC)。對于0.1PLC,增加0.005%量程補償指標,除了200mV、1A、10A量程,增加0.05%。對于0.01PLC,增加0.05%量程補償指標,除了200mV、1A、10A量程增加0.5%。
3.當正確歸零時,精度適用于2線或4線模式。
4.在脈沖模式,限制在0.1 PLC測量。
電流準確度(本地或遠端感測)
型號 |
量程 |
設置分辨率 |
電源 1, 3準確度(1 年)23°C ±5°C±(% rdg. + 電壓) |
缺省測量分辨率 |
測量 5, 6, 7準確度 (1 年)23°C ±5°C±(% rdg. + 電壓) |
源/阱極限 |
2400, 2401 |
1.00000 μA |
50 pA |
0.035% + 600 pA |
10 pA |
0.029% + 300 pA |
±1.05A @ ±21 V ±105 mA @ ±210 V 8 |
10.0000 μA |
500 pA |
0.033% + 2 nA |
100 pA |
0.027% + 700 pA |
||
100.000 μA |
5 nA |
0.031% + 20 nA |
1 nA |
0.025% + 6 nA |
||
1.00000 mA |
50 nA |
0.034% + 200 nA |
10 nA |
0.027% + 60 nA |
||
10.0000 mA |
500 nA |
0.045% + 2 μA |
100 nA |
0.035% + 600 nA |
||
100.000 mA |
5 μA |
0.066% + 20 μA |
1 μA |
0.055% + 6 μA |
||
1.00000 A 2 |
50 μA |
0.27 % + 900 μA |
10 μA |
0.22 % + 570 μA |
||
2410 |
1.00000 μA |
50 pA |
0.035% + 600 pA |
10 pA |
0.029% + 300 pA |
±1.05A @ ±21 V ±21 mA @ ±1100 V |
10.0000 μA |
500 pA |
0.033% + 2 nA |
100 pA |
0.027% + 700 pA |
||
100.000 μA |
5 nA |
0.031% + 20 nA |
1 nA |
0.025% + 6 nA |
||
1.00000 mA |
50 nA |
0.034% + 200 nA |
10 nA |
0.027% + 60 nA |
||
20.0000 mA |
500 nA |
0.045% + 4 μA |
100 nA |
0.035% + 1.2 μA |
||
100.000 mA |
5 μA |
0.066% + 20 μA |
1 μA |
0.055% + 6 μA |
||
1.00000 A 2 |
50 μA |
0.27 % + 900 μA |
10 μA |
0.22 % + 570 μA |
||
2420 |
10.0000 μA |
500 pA |
0.033% + 2 nA |
100 pA |
0.027% + 700 pA |
±3.15A @ ±21 V ±1.05 A @ ±63 V |
100.000 μA |
5 nA |
0.031% + 20 nA |
1 nA |
0.025% + 6 nA |
||
1.00000 mA |
50 nA |
0.034% + 200 nA |
10 nA |
0.027% + 60 nA |
||
10.0000 mA |
500 nA |
0.045% + 2 μA |
100 nA |
0.035% + 600 nA |
||
100.000 mA |
5 μA |
0.066% + 20 μA |
1 μA |
0.055% + 6 μA |
||
1.00000 A 2 |
50 μA |
0.067% + 900 μA |
10 μA |
0.066% + 570 μA |
||
3.00000 A 2 |
50 μA |
0.059% + 2.7 mA |
10 μA |
0.052% + 1.71 mA |
||
2440 |
10.0000 μA |
500 pA |
0.033% + 2 nA |
100 pA |
0.027% + 700 pA |
±5.25A @ ±10.5 V ±1.05 A @ ±42 V |
100.000 μA |
5 nA |
0.031% + 20 nA |
1 nA |
0.025% + 6 nA |
||
1.00000 mA |
50 nA |
0.034% + 200 nA |
10 nA |
0.027% + 60 nA |
||
10.0000 mA |
500 nA |
0.045% + 2 μA |
100 nA |
0.035% + 600 nA |
||
100.000 mA |
5 μA |
0.066% + 20 μA |
1 μA |
0.055% + 6 μA |
||
1.00000 A |
50 μA |
0.067% + 900 μA |
10 μA |
0.060% + 570 μA |
||
5.00000 A |
50 μA |
0.10 % + 5.4 mA |
10 μA |
0.10 % + 3.42 mA |
NOTES
- 溫度系數(0°C ~18°C & 28°C ~50°C):±(0.15×規定的準確度)/°C。
- 電流調整率:線路調整率:量程的0.01%。負載調整率:量程的0.01%(除了2440型 5A量程的0.05%) + 100pA。
- 電壓極限/嵌位:雙極性電壓極限(嵌位)設為一個值。量程的0.1%(最小值)。
- 過沖:典型值<0.1%(1mA步長,RL=10kΩ,20V量程,適于2400、2410、2420、2425 、2430),(10V量程,對于2440)。
注解:
1.僅2400、2410:指標對連續輸出電流低于105mA有效。對于高于105mA連續工作>1 分鐘,精度降低10%/35mA(高于105mA)。
2.全工作(1A)不考慮負載至30°C(50°C對于2420和2440)。高于30°C(50°C 對于2420和2440)環境溫度,降低35mA/°C并且分配35mA/Ω負載。4線模 式。對于電流阱工作在1A、3A或5A量程,最大連續功率限制在約1/2額定功率 或更低,這取決于電流,達30°C環境溫度。見用戶手冊的功率公式計算具體 條件下允許的占空比。
3.對于阱模式,1μA~100mA量程,準確度是: 2400:±(0.15%+偏移量*4),2410、2420、2425、2430、2440:±(0.5%+偏移 量*3)。
對于1A量程,準確度是: 2400:±(1.5%+偏移量*8),2410、2420、2425、2430、2440:±(1.5%+偏移量*3)
4.10A的量程僅在脈沖模式中可以達到。限制在2.5ms脈沖寬度最大值。10%占空比 最大值。
5.速度=正常(1PLC)。對于0.1PLC,增加0.005%量程補償指標,除了200mV、 1A、10A量程增加0.05%。對于0.01PLC,增加0.05%量程補償指標,除了 200mV、1A、10A量程增加0.5%。
6.當正確歸零時,精度適用于2線或4線模式。
7.在脈沖模式,限制在0.1 PLC測量。
8.僅2400和2400-C。
電阻測量準確度(本地或遠端感測) 1, 2, 5
量程 |
缺省分辨率 |
缺省測試電流 2400, 2401, 2410 |
缺省測試電流 |
標準準確度 (23°C ±5°C)1 年, ±(% rdg. + 電流) |
增強準確度(23°C ±5°C) 41 年, ±(%rdg.+電阻) |
||
2400, 2401 |
2410 |
2420, 2440 |
2400, 2401 |
||||
<0.20000 W 3 |
– |
– |
– |
Source I ACC + Meas. V ACC |
Source I ACC + Meas. V ACC |
Source I ACC + Meas. V ACC |
Source I ACC + Meas. V ACC |
2.00000 W 3 |
10 μW |
– |
1 A |
Source I ACC + Meas. V ACC |
Source I ACC + Meas. V ACC |
0.17% + 0.0003 W |
Source I ACC + Meas. V ACC |
20.0000 W |
100 μW |
100 mA |
100 mA |
0.10% + 0.003 W |
0.11% + 0.006 W |
0.10% + 0.003 W |
0.07% + 0.001 W |
200.000 W |
1 mW |
10 mA |
10 mA |
0.08% + 0.03 W |
0.09% + 0.1 W |
0.08% + 0.03 W |
0.05% + 0.01 W |
2.00000 kW |
10 mW |
1 mA |
1 mA |
0.07% + 0.3 W |
0.08% + 0.6 W |
0.07% + 0.3 W |
0.05% + 0.1 W |
20.0000 kW |
100 mW |
100 μA |
100 μA |
0.06% + 3 W |
0.07% + 6 W |
0.06% + 3 W |
0.04% + 1 W |
200.000 kW |
1 W |
10 μA |
10 μA |
0.07% + 30 W |
0.07% + 60 W |
0.07% + 30 W |
0.05% + 10 W |
2.00000 MW 6 |
10 W |
1 μA |
1 μA |
0.11% + 300 W |
0.12% + 600 W |
0.11% + 300 W |
0.05% + 100 W |
20.0000 MW 7 |
100 W |
1 μA |
1 μA |
0.11% + 1 kW |
0.12% + 2.4 kW |
0.11% + 1 kW |
0.05% + 500 W |
200.000 MW 3 |
1 kW |
100 nA |
– |
0.66% + 10 kW |
0.66% + 24 kW |
Source I ACC + Meas. V ACC |
0.35% + 5 kW |
>200.000 MW 3 |
– |
– |
– |
Source I ACC + Meas. V ACC |
Source I ACC + Meas. V ACC |
Source I ACC + Meas. V ACC |
Source I ACC + Meas. V ACC |
溫度系數(0°C ~18°C & 28°C ~50°C):±(0.15×規定的準 確度)/°C。
電流源模式,手動電阻:不確定度總和=電流源準確度+電壓測 量準確度(4線遠端感測)。
電壓源模式,手動電阻:不確定度總和=電壓源準確度+電流測 量準確度(4線遠端感測)。
6線電阻模式:支持使用有源電阻防護和防護感測。防護輸出
電流最大值:50mA(1A量程除外)。準確度依賴于負載。計算公式請 參見第2033號白皮書。
防護輸出阻抗:<0.1Ω,電阻模式。
注解
- 1.速度=正常(1 PLC)。對于0.1 PLC,增加0.005%量程補償指標,除了200mV、1A、10A量程增加 0.05%。對于0.01 PLC,增加0.05%量程補償指標,除了200mV、1A、10A量程增加0.5%。
- 2.當正確歸零時,精度適用于2線或4線模式。
- 3.僅手動電阻——除了2420、2425、2430、2440用于2Ω量程以及2410或2400用于200MΩ量程。
- 4.激活源回讀,偏移補償開。而且2410、2420、2425、2430和2440也提供類似的精度增強。
- 5.在脈沖模式,限制在0.1 PLC測量。
- 6.除了2440;缺省測試電流是5μA。
- 7.除了2440;缺省測試電流是0.5μA。
系統速度
測量 1
量程變化速率最大值:75/s
自動量程測量時間最大值:40ms(固定源)。
聯系人: | 魏成毅 |
---|---|
電話: | 0755-23357287 |
Email: | sales@falank.com |
QQ: | 3327351038 |
微信: | 18689488394 |
旺旺: | 法瀾克儀器設備 |
地址: | 廣東省深圳市寶安區石巖街道金三角商務大廈306 |